| Ex-MX2 Intrinsically Safe Infrared Thermometer with Laser Sighting
- For non-contact temperature measurement and monitoring of temperature processes in hazardous areas.
- Temperature range: from -22°F to +1652°F (-30 to +900°C)
- Quick response time: 250 ms
- Simple operation: trigger action
- Trend display: recall of previous 10 readings
- Reliable: laser circuitry provides measurement mark diameter
- Measurements: From Zone 1 to Zone 0
The intrinsically-safe non-contact temperature meter EX-MX2 is a robust, highly reliable and easy to use instrument for measuring the temperature of objects that are difficult to access, are moving or are as material on conveyor belts.
Operation could not be simpler: sight the object, press the trigger and the temperature can be read off the display immediately. The need to use the time consuming practice of focusing or calibration is now a thing of the past. The laser sighting allows exact areas - even on distant objects - to be pinpointed, so that high accuracy and an exact capture of a specific area can be guaranteed.
The Emissivity-Factor of 0.1 to 1.5 can be variably adjusted by increments of 0.01. This enables high measuring accuracy with only the emitted radiation part of a body being calculated from a cluster of emitted, reflected and transmitted radiation. Constant information about the actual temperature as well as minimum and maximum temperature values are displayed along with a graphical trend of the previous 10 readings and also the condition of the battery.
The back-lighted display enables readings to be taken even when used in dark environments and poor light conditions. The benefit of this simple yet robust device is that temperature monitoring as part of routine maintenance or manufacturing procedures can be easily carried out over a wide temperature range of -22 to + 1652°F (-30°C to +900°C) with an accuracy of + 1.8°F (1°C) (up to 212°F/100°C) and + 1% (above 212°F/100°C).
The EX-MX2 now allows high precision, repeat accuracy of temperature monitoring tasks to be carried out safely within the hazardous area.
Certifications: 
Certification:  II 2 G EEx ia IIC T4 Certificate of Conformity: TÜV 00 ATEX 1597 X
Documentation:
Certificate
Data Sheet
Manual
Specifications:
- Temperature range: -22°F to 1652°F (-30°C ...+900°C)
- Accuracy: (at ambient 68°F± 86°F)
- -22°F… 30.2°F ±3.6°F (-30°C … -1°C ±2°C)
- 32°F … 210.2°F ±1.8°F (0°C … 99°C ±1°C)
- 212°F … 1652°F ±1% ± 1.8°F or ± 1% (100°C … 900°C±1% ± 1°C or ± 1%) of reading - whichever is greater
- Repeatability: ±0.5% or ±1.8°F of reading whichever is greater
- Response time: (95%) 250 ms Indication elimination: 0.2°F (0.1°C) with temperature
- Emissivity adjustment: 0.1 to 1.5 (0.01 incremental)
- Display: actual temperature minimum temperature maximum temperature graphical trend (previous 10 readings) warning symbol emissivity level battery condition
- Ambient operating range: 32°F to 122°F (0°C … +50°C) - with laser max. +113°F
- Storage temperature: -4°F to +113°F (-20°C … +50°C)
- Relative humidity: 10-95% r.H. at up to 86°F
- Battery: IEC LR6/AA respectively R6/AA
- Dimensions: 7.8" x 6.6" x 2" (200 x 170 x 50 mm)
- Weight: approx. 16 oz (450 gr.)
- Spectral response: 8 to 14 µm
- Laser: class 2 laser
|