| Intrinsically-Safe Frequency Calibrator
For the measurement of revolutions as well as calibration and testing of flow-meters, rotary piston meters, turbine meters and oval gear meters etc. in the hazardous area.
Operating mode
- Predetermining / measuring frequency
- Predetermining / measuring pulses
- Measuring/simulating number of revolutions
- Testing NAMUR-initiators
- Predetermining/measuring times
- Transmitting/measuring static levels
The FCX2000 offers a simple and quick means of measuring and simulating frequencies, pulses, numbers of revolutions, digital process signals and levels. The operator guidance with the help of a 4- Iine LCD-display and membrane keyboard allows simple and easy to use adjustment of all operator functions.
lt is possible to predetermine and measure NAMUR- signals according to DIN 19234 as well as static logic levels in hazardous locations, ranging from 0.01 Hz to 10 kHz with duty factors from 1 to 99 %. Accurate numbers of revolutions can either be measured by NAMUR initiators or by an optional external probe. For measuring revolutions on rotating machinery with the probe a suitable reflection mark needs to be attached. A visible beam of red light from the probe is then trained on this reflective mark.
An important function for flow-metering is the FCX2000 in the mode "pulse pre-determining" or "pulse measuring". Here pulse length and pulse pause will be predetermined first, before a fixed programmed number of pulses is released. The display in this mode shows, what percentage of the programmed pulses have been sent. Whether a two conductor NAMUR-initiator works properly or is malfunctioning can be checked by another function.
The intrinsically safe FCX2000 is supplied with standard rechargeable batteries. These are easily removed, but any recharging must be done outside the hazardous area. An integral battery level indicator shows when the batteries are becoming exhausted, so that enough reserve operating time is ensured.
The lightweight metal housings designed to ensure that the electronic parts of the FCX2000 are well protected so that the equipment can be used safely and with confidence under the harshest of industrial conditions.
Certifications:
Certification:  EEx ia IIC T6  EEx ib IIC T6 Certificate of Conformity: PTB-Nr. Ex-92.C.2116  FM Approved Class I Zone 0 AEx ia IIC T3 Class I Divison 1 Group A-D
Documentation:
Certificate
Data Sheet
Manual
T3C FCX 2000 as transmitter:
- Umax = 10,5V
- IK = 14,1mA P = 74mW
- EEx ia IIC
- EEx ib IIC
FCX 2000 as a measuring device:
Specifications:
- Display: 4-lines with 16 alphanumeric symbols each
- Discharge indicator: red LED
- Operation temperature: -10°C ... +40°C
- Storage temperature: -10°C ... +60°C
- Reference temperature: 20°C ± 3 K
- Air moisture: 0...90% r. H. of the air
- Supply voltage internal:
- 6 x standard batteries AA LR6/R6 according to IEC
- 6x rechargeable batteries
- Typ 5006, VARTA
- Typ VE AA 700, SAFT
- Battery change: outside hazardous area
- Operation time: 9 hours - continuous
- Inputs: protected against unintended reversing of polarity
- Outputs: short curcuit protected Ingress protection: IP 54
- Housing material: aluminium
- Dimensions: 185 x 105 x 45 mm (without probe)
- Weight: approx. 1000g
|